Name/Function

Autoclave

Manufacture type

Getinge GE-406

Manufacture

Getinge

Made in

Sweden

Part number.

81136000

Production year

1982

Comment's

Needs a connection to water pipe and outflow.


 
Contents list: Autoclave 81136000

Part number

Description

Qty
Note

 

Contents list

1

 

Edition 1 1982

Instruction manual for Autoclave GE-406

1

 

 

Test certificate from the government inspection unit

1

 

 

 

 

 

 

Autoclave Getinge GE-406

1

 

 

 

 

 

 

Upper grating

1

 

 

Lower grating

1

 

 

 

 

 

460 40 21-48

Graph paper pack of 100 pcs

1

 

 

 

 

 

470 26 96 -01

Tools for disassembling pumps

1

 


Technical data

Width cabinet type

750 mm

Steam consumption
 

Width recessed type

1050 mm

/process(135 textils progr)

4 kg

Height cabinet type

1660 mm

peak value

30kg/h

Height recessed type

1700 mm

Pilot voltage

24 VAC

Depth

930 mm

Fuse pilot circuits 5 x 20 mm

10 A

Loading height

800 mm

Electricity alternative 1

 

Weight

360 kg

3 x 380 volts + Beutral, 50 Hz

 

Chamber width

455 mm

Electricity alternative 2

 

heigt

460 mm

3 x 220 volts 50 Hz

21 kW

depth

660 mm

Elecric rating incl. steamgenerator

0,9

corner radius

75 mm

Elecric rating exkl. steamgenerator

kW

volume

135 litre

Wlectric energy consumption/process

 

design pressure

300 kpa

Incl. steam generator

3,6 kWh

material SIS 2347

 

Steam generator

 

corresponding to AISI 316

 

Gross volume

21 litre

Vacuumpump electric rating

0,75 kW

Water volume

10 litre

Water consumption at 110 mbar

3/l/min

Design pressure

350 kpa

Feedwater pump electric rating

0,37 kW

Heat losses

0,5 kW

Feedwater pumpcapacity at 400 kpa

 

Electric rating

19 kW

Conterpressure

5,6 l/min

 

 

Heat losses,stream generator

 

 

 

equipped

about 2 kW

 

 

Water consumption/process

200 litre

 

 

Peak value

10 l/min

 

 

 

 

 

 


Skapad med publiceringsverktyget eContent.